Programmable Temperature Controller (Digital Controller)
E5CN-HT
(48 x 48 mm)
CSM_E5CN-HT_DS_E_1_9
Programmable Controllers Join the
Series!
Program up to 256 segments and take
advantage of the high cost performance
of the new LCD that improves both the
field of view and contrast.
• Set up to 8 programmed patterns with up to 32 segments (steps)
each.
• High-resolution display with 5 digits/0.01°C display in a compact
Controller (48 x 48 mm).
• High-speed sampling cycle of 60 ms.
• High Accuracy
Thermocouple/Pt input: ±0.1% of PV
Analog input: ±0.1% FS
• Universal inputs on all models (thermocouple, PT, or analog
input) to handle various sensors with one Controller.
• A PV/SV-status display function can be set to automatically alternate
between displaying the status of the Temperature Controller (auto/
manual, RUN/RESET, and alarms) and the PV or SV.
• Flexible contact outputs with logic operations (AND, OR, and
delays) set from the Support Software (CX-Thermo version 4.3).
Program settings can be managed.
• Preventive maintenance for relays in the Temperature Controller
using a Control Output ON/OFF Counter.
48
×
48 mm
E5CN-HT
For the most recent information on models that have been certified
for safety standards, refer to your OMRON website.
Refer to
Safety Precautions for E5
@
N/E5
@
N-H/
E5
@
N-HT.
Refer to
Operation for E5
@
N/E5
@
N-H/ E5
@
N-HT
for
operating procedures.
E53-CNQ@-500 in this catalog have been
discontinued at the end of March 2018.
Main I/O Functions
Event Inputs
•
None
•
Two
Universal Sensor Inputs
•
Universal thermocouple/Pt/analog
(current/voltage) inputs.
Indication Accuracy
•
Thermocouple/Pt input:
±0.1%
of PV
•
Analog input:
±0.1%
FS
Sampling Period
•
60 ms
Control Output 2
*
•
None
•
Voltage output
(for driving SSR)
Control Output 1
•
Relay output
•
Voltage output (for driving SSR)
•
Current output
•
Linear voltage output
E5CN-HT
2-level Display: PV and SV
5-digit Display with high resolution of 0.01°C
•
Programmable function
8 programs × 32 segments
•
Auto/manual switching
•
Temperature Controller status display
•
Simple program function
•
Logic operations
•
Control output ON/OFF count alarm
•
PV change rate alarm
•
Models also available with RS-485 or
RS-232C communications
Auxiliary Outputs
•
Two
*
Orders will not be accepted after
March 31, 2018.
This data sheet is provided as a guideline for selecting products. Be sure to refer to the following user manuals for application precautions
and other information required for operation before attempting to use the product.
E5CN-HT/E5AN-HT/E5EN-HT Digital Controllers User's Manual Programmable Type (Cat. No. H169)
E5CN-HT/E5AN-HT/E5EN-HT Digital Controllers Communications Manual Programmable Type (Cat. No. H170)
1
E5CN-HT
Lineup
E5CN-HT
Programmable
Type
1 control output
Terminal block
Fully universal input
2 control outputs
2 auxiliary outputs
2 auxiliary outputs
Note:
Models with one control output and models with two control outputs can be used for heating/cooling control.
Model Number Structure
Model Number Legend
Controllers
Option Units
E5CN-@@@M@-@-500
1 2 3 4 5
1. Type
HT: Programmable
2. Control Output 1
R: Relay output
Q: Voltage output (for driving SSR)
C: Current output
V: Linear voltage output
3. Auxiliary Outputs
2: Two outputs
4. Option 1
M: Option Unit can be mounted.
5. Power Supply Voltage
Blank: 100 to 240 VAC
D: 24 VAC/VDC
6. Case Color
Blank: Black
7. Terminal Cover
-500: With terminal cover
E53-@@@@
1 2 3 4
1. Applicable Controller
CN: E5CN-HT, E5CN-H or E5CN
2. Function 1
Blank: None
Q: Control output 2 (voltage output for driving SSR)
*
P: Power supply for sensor
C: Current output
3. Function 2
Blank: None
H: Heater burnout/SSR failure/Heater overcurrent detection (CT1)
HH: Heater burnout/SSR failure/Heater overcurrent detection
(CT2)
B: Two event inputs
03: RS-485 communications
H03: Heater burnout/SSR failure/Heater overcurrent detection
(CT1) + RS-485 communications
HB: Heater burnout/SSR failure/Heater overcurrent detection
(CT1) + Two event inputs
HH03: Heater burnout/SSR failure/Heater overcurrent detection
(CT2) + RS-485 communications
H01: Heater burnout/SSR failure/Heater overcurrent detection
(CT1)/RS-232C communications
F: Transfer output
BF: Two event inputs/Transfer output
4. Version
N2: Available only to models released after January 2008
6
7
Note: 1.
Not all combinations of function 1 and function 2 specifications are possible for Option Units (E53-@@@@).
2.
Estimates can be provided for coatings and other specifications that are not given in the datasheet. Ask your OMRON representative for details.
*
Orders will not be accepted after March 31, 2018.
2
E5CN-HT
Ordering Information
Controllers
Size
Case Color
Power supply voltage
Auxiliary output
Control output 1
Relay output
100 to 240 VAC
1/16 DIN
48
×
48
×
78
(W
×
H
×
D)
Black
24 VAC/VDC
2
2
Voltage output (for driving SSR)
Current output
Linear voltage output
Relay output
Voltage output (for driving SSR)
Current output
Linear voltage output
Model
E5CN-HTR2M-500
E5CN-HTQ2M-500
E5CN-HTC2M-500
E5CN-HTV2M-500
E5CN-HTR2MD-500
E5CN-HTQ2MD-500
E5CN-HTC2MD-500
E5CN-HTV2MD-500
Option Units
One of the following Option Units can be mounted to provide the E5CN with additional functions.
Functions
Communications
RS-485
3-phase heater burnout/SSR failure/
Heater overcurrent detection
Heater burnout/SSR failure/
Heater overcurrent detection
Communications
RS-485
Communications
RS-485
Communications
RS-485
Event inputs
Heater burnout/SSR failure/
Heater overcurrent detection
3-phase heater burnout/SSR failure/
Heater overcurrent detection
Event inputs
Control output 2
(Voltage for driving SSR)
Control output 2
(Voltage for driving SSR)
Control output 2
(Voltage for driving SSR)
Control output 2
Transfer Output
(Voltage for driving SSR)
Event inputs
Communications
RS-232C
Communications
RS-232C
Communications
RS-232C
Heater burnout/SSR failure/
Heater overcurrent detection
Control output 2
(Voltage for driving SSR)
Transfer Output
Heater burnout/SSR failure/
Heater overcurrent detection
Event inputs
Control output 2
(Voltage for driving SSR)
Model
E53-CNHH03N2
E53-CNHBN2
E53-CNQ03N2
*
E53-CNH03N2
E53-CN03N2
E53-CNBN2
E53-CNQHN2
*
E53-CNQHHN2
*
E53-CNQBN2
*
E53-CNQFN2
*
E53-CNBFN2
E53-CNQ01N2
*
E53-CN01N2
E53-CNH01N2
Note:
These Option Units are applicable only to models released after January 2008.
*
Orders will not be accepted after March 31, 2018.
Accessories (Order Separately)
USB-Serial Conversion Cable
Model
E58-CIFQ1
Adapter
Connectable models
Terminal type
Model
Y92F-45
Terminal Cover
Model
E53-COV17
Note:
Use this Adapter when the panel has been previously prepared
for the E5B@.
DIN Track Mounting Adapter
Model
Y92F-52
Note: 1.
The Terminal Cover comes with the E5CN-@@@-500
models.
2.
The E53-COV10 cannot be used.
Waterproof Packing
Model
Y92S-P8
Front cover
Type
Hard Front Cover
Soft Front Cover
Model
Y92A-48B
Y92A-48D
Note:
Waterproof Packing is included with the controller only for
models with terminal blocks.
CX-Thermo Support Software
Model
EST2-2C-MV4
Current Transformers (CTs)
Hole diameter
5.8 dia.
12.0 dia.
Model
E54-CT1
E54-CT3
3
E5CN-HT
Specifications
Ratings
Power supply voltage
Operating voltage range
Power consumption
No D in model number: 100 to 240 VAC, 50/60 Hz
D in model number: 24 VAC, 50/60 Hz; 24 VDC
85% to 110% of rated supply voltage
100 to 240 VAC: 8.5 VA (max.) (E5CN-HTR2 at 100 VAC: 3.0 VA)
24 VAC/VDC: 5.5 VA (24 VAC)/3.5 W (24 VDC) (max.) (E5CN-HTR2D at 24 VAC: 2.7 VA)
Any of the following can be selected (i.e., fully universal input).
Thermocouple: K, J, T, E, L, U, N, R, S, B, W, or PL II
Platinum resistance thermometer: Pt100 or JPt100
Current input: 4 to 20 mA or 0 to 20 mA
Voltage input: 1 to 5 V, 0 to 5 V, or 0 to 10 V
Current input: 150
Ω
max., Voltage input: 1 MΩ min. (Use a 1:1 connection when connecting the ES2-HB-N.)
ON/OFF control or 2-PID control (with auto-tuning)
SPST-NO, 250 VAC, 3 A (resistive load), electrical life: 100,000 operations, minimum applicable load: 5 V, 10 mA
4 to 20 mA DC/0 to 20 mA DC, load: 600
Ω
max., resolution: approx. 10,000
*
0 to 10 VDC (load: 1 kΩ min.), Resolution: Approx. 10,000
Sensor input
Input impedance
Control method
Relay output
Control
output
Current output
Linear voltage
output
Number of outputs
2 max.
Auxiliary
Output
Relay output: SPST-NO, 250 VAC, 3 A (resistive load), electrical life: 100,000 operations, minimum applicable
output
specifications
load: 5 V, 10 mA
Number of outputs
2
Event
input
External contact
input
specifications
Number of
operations
Contact input: ON: 1 kΩ max., OFF: 100 kΩ min.
Non-contact input: ON: Residual voltage: 1.5 V max., OFF: Leakage current: 0.1 mA max.
Current flow: Approx. 7 mA per contact
8 max. (Combinations can be made using work bits.)
•
Logic operation: Any of the following four patterns can be selected. The input status may be inverted.
(A and B) or (C and D), (A or C) and (B or D), A or B or C or D, A and B and C and D (A, B,
C, and D are four inputs.)
•
Delay:
ON delay or OFF delay for the results of the logic operation given above.
Setting time: 0 to 9999 s or 0 to 9999 min
•
Output inversion: Possible
One work bit per operation
Any of the following can be assigned to up to eight work bits (logic operation results): Operation commands
(assigned to event inputs)
*,
auxiliary outputs, or control outputs.
Logic
opera-
tions
Operations
Outputs
Work bit
assignments
*
Application is possible with models that do not have event inputs by using an internal assignment.
Number of outputs
1 max.
Transfer
outputs Output
Current output: 4 to 20 mA DC, Load: 600
Ω
max., Resolution at 4 to 20 mA: Approx. 10,000
specifications
RSP input
Setting method
Indication method
Not supported
Digital setting using front panel keys
11-segment digital display and individual indicators (7-segments displays also possible)
Character height: PV: 11 mm, SV: 6.5 mm
Manual output, heating/cooling control, loop burnout alarm, other alarm functions, heater burnout detection
(including SSR failure and heater overcurrent detection), 40% AT, 100% AT, MV limiter, input digital filter,
temperature input shift, run/reset, protection functions, control output ON/OFF counter, extraction of square root,
MV change rate limit, PV/SV status display, automatic cooling coefficient adjustment, program control functions,
etc.
−10
to 55°C (with no condensation or icing), for 3-year warranty:
−10
to 50°C
25% to 85%
Other functions
Ambient operating
temperature
Ambient operating humidity
Storage temperature
−25
to 65°C (with no condensation or icing)
*
For models with current outputs, control output 1 can be used as a transfer output.
4
E5CN-HT
Input Ranges
Thermocouple/Platinum Resistance Thermometer/Analog Input (Fully Universal Inputs)
Input type
Name
2300
1800
1700
1600
1500
1400
1300
1200
1100
1000
900
800
700
600
500
400
300
200
100
0
−100
−200
−200.0−199.9
200.00
200.00
Platinum resistance
thermometer
Pt100
JPt100
K
J
T
Thermocouple
E
L
U
N
R
S
B
1800.0
1700.01700.0
Analog
input
W
PL II
2300.0
4 to
0 to
20 mA 20 mA
1 to
5V
0 to 0 to
5 V 10 V
1300.0
1300.0
1300.0
Temperature range (°C)
850.0
850.0
850.0
600.0
500.0
500.0
500.0
400.0
200.00
Usable in the following
ranges by scaling:
−19999
to 32400,
−1999.9
to 3240.0,
−199.99
to 324.00, or
−19.999
to 32.400
400.0 400.0
400.0 400.0
200.00
100.0
0.0
−50.00
−199.9
100.0
100.0
0.0
−20.0 −50.00−100.0 −20.0 −50.00
−200.0
−200.0−199.9
−50.00
−200.0
−100.0
−200.0−199.9−200.0
0.0
0.0
0.0
0.0
Setting
number
0
1
2
24
3
4
5
6
21
7
8
22
9
10
23
11
12
13
14
15
16
17
18
19
20
25
26
27
28
29
Shaded settings are the default settings.
The applicable standards for the input types are as follows:
K, J, T, E, N, R, S, B: JIS C 1602-1995, IEC 584-1
L: Fe-CuNi, DIN 43710-1985
U: Cu-CuNi, DIN 43710-1985
W: W5Re/W26Re, ASTM E988-1990
JPt100: JIS C 1604-1989, JIS C 1606-1989
Pt100: JIS C 1604-1997, IEC 751
PL II: According to Platinel II electromotive force charts from BASF (previously
Engelhard)
5